Abstract

A new assay method has been developed for the quantitative determination of heparin in serum using a microtiter plate-format optode (MPO). Heparin and proton in physiological sample are favorably co-extracted into the solvent polymeric optode membrane containing both cationic lipophilic additive, tridodecylmethyl ammonium chloride (TDMAC), and proton-selective ionophore, 3-hydroxy-4-(4-nitrophenylazo)-phenyloctadecanoate (ETH 2412), resulting in the absorbance change of the membrane to varying heparin levels. The optimized MPO composition contains low polymer-to-plasticizer ratio compared to those of conventional ion-selective optodes or electrodes, i.e., poly(vinyl chloride) (20.0)/dioctylsebacate (76.3)/ETH 2412 (1.7)/TDMAC (1.0) (wt.%): it resulted in a quantitative response to heparin from 0 to 15 unit/mL in serum with high sensitivity. The heparin–protamine titration on the MPO could provide rapid and precise determination of heparin. It was shown that the heparin levels in serum sample could be determined from the rate of absorbance change over time (Δ A/Δ t); this method was more effective than the direct absorbance measurement in minimizing the interferences from color and turbidity of serum samples. MPO has been developed as a high throughput and convenient disposable sensing device, and may find a wide application in the determination of polyions and charged macromolecules.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call