Abstract

Within the scope of this ex vivo study, the cut efficiency was investigated with eight diode laser wavelengths in the range from 400 to 1500 nm. Incisions on porcine gingiva samples were generated in CW-mode at a power range of 0.5–4 W using a bare fiber (∅ = 320 μm) in contact and non-contact mode at a cut speed of 2 mm/s. Cut depths, cut widths, and thermal damages were recorded based on histological sections and were evaluated via measurement masks. Moreover, with respect to the controllability of a therapeutic measure, an efficiency factor was defined. At powers above 2 W, for 445 nm, the maximum cut depth was 820 μm and 344 μm for 810 nm, respectively. At all wavelength and power ranges, the cut width averaged 125 μm. At minimum output power (0.5 W), the spatial expansion of the thermal damage in the tissue surface layer corresponds in the blue/green wavelength range from the very beginning of the laser impact to the fiber core diameter. It could be shown that increases in the diode laser power output do not correlate to the same extent with the incision depth nor with thermal damage to tissue.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.