Abstract
A selective, rapid and sensitive ultra performance liquid chromatography–tandem mass spectrometry (UPLC–MS/MS) method was developed for the quantitative determination of azithromycin in human plasma and its application in a pharmacokinetic study. With roxithromycin as internal standard, sample pretreatment involved a one-step extraction with diethyl ether of 0.5 mL plasma. The analysis was carried out on an ACQUITY UPLC™ BEH C 18 column (50 mm × 2.1 mm, i.d., 1.7 μm) with gradient elution at flow rate of 0.35 mL/min. The mobile phase was 50 mM ammonium acetate and acetonitrile. The detection was performed on a triple-quadrupole tandem mass spectrometer by multiple reaction monitoring (MRM) mode via electrospray ionization (ESI). Linear calibration curves were obtained in the concentration range of 1–1000 ng/mL, with a lower limit of quantification of 1 ng/mL. The intra- and inter-day precision (RSD) values were below 15% and accuracy (RE) was −1.3% to 5.7% at all QC levels. The method was applicable to clinical pharmacokinetic study of azithromycin in healthy volunteers following oral administration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.