Abstract

A rapid, sensitive, and selective ultra fast liquid chromatography-tandem mass spectrometry method was developed for quantitative determination of arenobufagin in rat plasma. Sample pretreatment involved a one-step protein precipitation with methanol using 0.1mL rat plasma. The separation was carried out on a Shim-pack XR-ODS II (75mm×2.0mm, i.d. 2.1μm) column with gradient elution at a flow rate of 0.30mLmin(-1). The mobile phase was acetonitrile and 0.1% formic acid in water. A post-column switching valve was applied to reduce the matrix effect. The detection was performed on a triple-quadruple tandem mass spectrometer in the multiple reaction monitoring mode after electrospray ionization. Linear calibration curves for arenobufagin were obtained over the concentration range 1.056-1056ngmL(-1), with a lower limit of quantification of 1.056ngmL(-1). The intra-day and inter-day precision values were lower than 15% and the accuracy ranged from 5.4% to 9.8% at all quality control levels. The method was successfully applied to the determination and pharmacokinetic study of arenobufagin in rat plasma following intraperitoneal administration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.