Abstract

A simple isocratic and stability-indicating HPLC method was developed and validated for the quantitative determination of anti-tumor agent fluorapacin and its pharmaceutical preparation. A Spherisorb ODS II C 18 (250 mm × 4.6 mm, 5 μm) column was eluted with a mobile phase consisting of acetonitrile/water (85:15, v/v). The analyses were performed at 40 ± 1 °C with a flow rate of 1.0 mL/min and UV detection at 218 nm. The calibration curve was linear over a concentration range of 160–240 μg/mL with the correlation coefficient of 0.9997. The LOD and LOQ were determined to be 1.4 and 7.0 ng/mL, respectively. Average recoveries were 98.27% and 100.40% for fluorapacin API and its drug product with corresponding relative standard deviations (R.S.D.) of 0.41% and 0.30%, respectively. Good repeatability (precision and intermediate precision), accuracy and tolerability were obtained with R.S.D. of <1.0%. This specific and reliable method has been successfully applied for quality control of fluorapacin API and drug product.