Abstract

A corona-charged aerosol detector (CAD) was developed to improve the sensitivity, reproducibility and quantitativeness of detection as compared to evaporative light-scattering detector (ELSD) for liquid chromatography. Our laboratory used the corona CAD as a detector for supercritical fluid chromatography (SFC) and evaluated its performance compared to the ELSD by using a certified reference material of poly(ethylene glycol) (PEG) and a well-defined equimass mixture of uniform PEG oligomers. The corona CAD was able to detect a 10 times more dilute solution of uniform oligomers compared to the ELSD. Although the original data of molecular mass by ELSD was 4.6% smaller than the certified value of PEG 1000, molecular mass distribution obtained by corona CAD was virtually almost the same as the certified value without any calibrations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.