Abstract

A corona-charged aerosol detector (CAD) was developed to improve the sensitivity, reproducibility and quantitativeness of detection as compared to evaporative light-scattering detector (ELSD) for liquid chromatography. Our laboratory used the corona CAD as a detector for supercritical fluid chromatography (SFC) and evaluated its performance compared to the ELSD by using a certified reference material of poly(ethylene glycol) (PEG) and a well-defined equimass mixture of uniform PEG oligomers. The corona CAD was able to detect a 10 times more dilute solution of uniform oligomers compared to the ELSD. Although the original data of molecular mass by ELSD was 4.6% smaller than the certified value of PEG 1000, molecular mass distribution obtained by corona CAD was virtually almost the same as the certified value without any calibrations.