Abstract

We investigated the use of the Schlieren method to calibrate the sound pressure in an ultrasonic standing-wave field. Specifically, we derived an equation to calculate the light intensity of the diffraction fringe induced by the standing-wave field. The results indicated that the sound pressure in the standing-wave field relates to the light intensity of the diffraction fringe. Simulations and experiments were conducted to verify the theoretical calculation. We demonstrated that the ratio of the light intensity of different diffraction orders relates to the sound pressure amplitude, allowing the pressure amplitude to be calibrated with the Schlieren method. Therefore, this work presents a non-intrusive calibration method that is particularly suitable for calibrating high-frequency ultrasonic standing-wave fields.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.