Abstract
The clinical applicability of a method for quantifying pericardial effusion (PE) was studied. The pericardial and cardiac volumes were determined from border tracings of 2-dimensional echocardiograms fitted with 3-dimensional disk models. The PE volume was the difference between pericardial and cardiac volumes. A phantom study included 54 cases with different volumes and viewing geometry showed a correlation coefficient r = 0.98 and an accuracy of ± 6%. A clinical study included 20 cases showing that the estimated PE volume correlated well with that drained surgically (between 100 mL and 1200 mL): y = 0.81 × + 120 mL; r = 0.91, P <.0001. The percent error, determined by the standard error of the estimate (114 mL) over mean (548 mL), was 20%. Intraobserver variability was 4% and interobserver variability 6%. The 3-dimensional disk method provides a quick and convenient way to quantify PE from 2-dimensional echocardiograms with acceptable accuracy and reproducibility. (J Am Soc Echocardiogr 2003;16:147-53.)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Society of Echocardiography
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.