Abstract

The analytical possibilities of quantification of the intact monoclonal antibody trastuzumab by high-performance liquid chromatography coupled with electrospray mass spectrometry (HPLC–ESI-MS) were investigated. To clarify the results obtained by LC–MS, complementary experiments were performed using direct UV-spectrophotometry and high-performance liquid chromatography coupled with ultraviolet detection (HPLC–UV). A polystyrene–divinylbenzene (POROS) column was applied with gradient elution using formic acid 0.08% (v/v) in water and formic acid 0.08% (v/v) in acetonitrile as mobile phase for chromatographic analysis. Quantification on LC–MS was performed by using the peak area of the total ion current (TIC) chromatograms of one charge state. Non-linearity and sensitivity loss were the major limitations observed with the LC–MS method, of which the non-linearity is most likely caused by detector saturation. The sensitivity loss during analysis could be reduced by lowering the MS source temperature. This parameter is critical in creating a robust LC–MS system for the quantitative analysis of trastuzumab.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.