Abstract
Neonicotinoids, including imidacloprid, are extensively used for plant protection against insects. Unfortunately, these effective pesticides are one of the reasons for the decline of the bee population in recent decades. Ensuring application of minimal pesticide quantity and preventing excess requires a fast method for monitoring the coverage on plants. We present a new, method based on multiphoton electron extraction spectroscopy (MEES), for detecting, quantifying and mapping of imidacloprid coverage on plants. Imaging and quantitative analyses were demonstrated on several plant surfaces including olive and mint leaves and orange peel. Method provides both low detection limits (down to nanogram level) and good trueness. This method is fast and can be directly performed with no sample pre-treatment, thus, it is a good candidate for field analyses.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.