Abstract

A simple, rapid and sensitive LC–MS/MS method in positive ion mode was developed and validated to determine CKD-501, lobeglitazone, in human plasma and urine using glipizide as an internal standard (IS). Lobeglitazone is a novel thiazolidinedione (TZDs)-based peroxisome proliferator-activated receptor (PPAR) agonist, used for the management of type-2 diabetes. After mixing the IS, dissolved in acetonitrile, with a plasma or urine sample containing lobeglitazone, 10 μL of supernatant was injected into the LC–MS/MS system. Quantification was performed in the multiple reaction monitoring (MRM) mode using transition of 481.5 → 152.2 (m/z) for lobeglitazone and 446.1 → 321.2 (m/z) for the IS. The method showed good linearity over concentration ranges of 0.5–1,000 ng mL−1 for plasma and 0.2–250 ng mL−1 for urine (r 2 ≥ 0.9996). The mean percent extraction recovery of lobeglitazone was 90.8 % for plasma and 87.3 % for urine, while the recoveries of the IS were greater than 86.4 % for both. The intra-day and inter-day precision of plasma ranged from 1.1 to 3.7 and 2.5 to 3.3 % (RSD), respectively, and the intra- and inter-day precision of urine ranged from 1.5 to 2.7 and 3.2 to 3.5 %, respectively. This method is simple, sensitive, and applicable for the pharmacokinetic study of lobeglitazone in human plasma. Most of the urine concentrations of lobeglitazone were below the LLOQ because the lobeglitazone is extensively metabolized.