Abstract
In this paper, a two-detector measuring system in in-air PIXE system composed of two Si(Li) detectors has been developed for simultaneous measurement of low- and high-Z elements. In order to improve detection sensitivity of the detector for low energy region, a new device which is attached at the tip of the detector has been designed. It is made of acryl and has a thin end on which a 1.5 μm-thick Mylar film is stuck. As a result, it exhibited a miraculous effect in improving detection sensitivity at low energies and it became possible to detect K X-rays of aluminium. In order to perform quantitative analysis in in-air system, we have measured detection efficiencies for the two Si(Li) detectors including the effect of X-ray absorption in air on the basis of the method that we developed. Concerning the beam energy at the target and corresponding X-ray production cross-sections, the same values as were reported in the previous paper were applicable since conditions of irradiating system were unchanged. It was confirmed that the new method allows us to quantitatively analyze all the elements heavier than aluminum and to obtain mostly the same results as those by in-vacuum PIXE for various kinds of samples. Accuracy of analysis was also confirmed by using a standard material.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.