Abstract

The cytotoxic drug cyclopentenyl cytosine (CPEC) is currently being investigated in early clinical trials. Monitoring of plasma levels is required for pharmacokinetic analysis and management of toxicity. This paper describes the analysis of CPEC and cyclopentenyl uracil (CPEU) in plasma by HPLC-electrospray ionization tandem mass spectrometry. The calibration curves for CPEC and the metabolite CPEU were linear up to 10 microm with correlation coefficients of 0.997 (SD = 0.002, n = 10) and 0.997 (SD = 0.004, n = 10), respectively. The detection limit for CPEC was 0.03 and 0.12 microM for CPEU. The intra- and interassay coefficients of variation for CPEC and CPEU were less then 10%. The usefulness of the method was demonstrated by analyzing CPEC and CPEU in plasma of a patient treated with CPEC. HPLC with electrospray ionization tandem mass spectrometry allowed rapid and sensitive determination of CPEC and CPEU levels in plasma.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.