Abstract

This study demonstrates rapid quantitation of ractopamine in beef using automated online sample preparation with turbulent-flow chromatography technology coupled to a triple stage quadrupole mass spectrometer. The total sample preparation time is approximately 30 minutes and the LC/MS method length is less than 6 minutes, resulting in a considerable reduction in overall analysis time over current methodologies. The assay has a linear dynamic range of 0.3–120 ng g−1 for ractopamine in beef. The limit of quantitation (LOQ) was determined at 0.3 ng g−1 using 25 μL injection volume. The recoveries were in the range of 85–104% at 0.6–60 ng g−1. The present approach is suitable for routine ractopamine residual monitoring in food testing laboratories.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.