Abstract
Raman analyses of fluid inclusions can yield quantitative information on composition (from peak areas and heights) and density (from peak position and width). In this study, we examine the effect of instrumental spectral resolution on the ratios of these spectral parameters, and the selection of appropriate integration limits for the determination of peak areas in the CO2-CH4-N2 system. Spectral resolution was varied from about 1 to 9 cm−1 by co-varying the widths of all spectrometer slits. Changes in resolution produced a modest effect on peak-area ratios and a significant effect on peak-height ratios. Measured peak-width ratios varied strongly as a function of the spectral resolution. In addition, we observed a moderate shift in the measured peak position of N2, which can be related to the asymmetry of the band. These results indicate that accurate analysis requires careful attention to the selection of quantification factors, especially if the selected values were derived from studies at different spectral resolutions. Another factor that can have a significant effect on the calculated compositions of CH4- and H2-bearing fluid mixtures is the band broadening that occurs with increasing pressure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.