Abstract

Anhydrous hydrogen chloride at concentrations of up to 100% has been introduced into a sealed inductively coupled plasma system for quantitative spectrochemical analysis. Vapour-phase sampling of monobutyltin trichloride was developed to calibrate tin and carbon. The addition of chlorine as a modifier gas was required in a ratio of 1:1 to maintain stable and reproducible emission signals. Under flowing conditions detection limits for tin and carbon were 49 and 271 (ng g–1), respectively, in a 16% v/v HCl–Cl2 argon plasma. Impurities identified qualitatively included Al, C, Ca, Cr, Fe, Ni and Sn. In addition, the bulk temperature of the plasma was determined to be 10 500 ± 1500 K.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.