Abstract

We present a quantitative clinical LC-MS/MS assay for the simultaneous analysis of buprenorphine, norbuprenorphine, and their glucuronide metabolites in human urine. The assay is based on the direct and hydrolysis-free sample preparation protocol, i.e., dilute and shoot, whereby clarified urine specimens are diluted in internal standard reagent and injected into the LC-MS/MS instrument. The analytical platform employs reversed-phase liquid chromatography for upfront separation and electrospray ionization multiple reaction monitoring MS detection via the triple-quadrupole (TSQ Quantiva) instrument. The assay has a quantitative analytical range of 5ng/mL-1000ng/mL represented at seven levels for each of the four analytes. A unique stable isotopically labeled analogue is used as internal standard for each analyte. For high-throughput performance, the assay can be multiplexed between two LC channels.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call