Abstract

Determination of quantitative compositions of blended oils is an essential but challenging step for the quality control and safety assurance of blended oils. We herein report a method for the quantitative analysis of blended oils based on the intensity ratio of triacylglycerol marker ions, which could be obtained from the highly reproducible spectra acquired by using matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) to directly analyze blended oils in their oily states. We demonstrated that this method could provide good quantitative results to binary, ternary, and quaternary blended oils, with simultaneous quantitation of multiple compositions, and was applicable for quantitative analysis of commercial blended oil products. Moreover, the intensity ratio-based method could be used to rapidly measure the proportions of oil compositions in blended oils, only based on the spectra of the blended oils and related pure oils, making the method as a high-throughput approach to meet the sharply growing analytical demands of blended oils.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.