Abstract
An ion mobility spectrometer (IMS) equipped with a 10.6 eV low-pressure gas-discharge lamp usually used in photoionization detectors for gas chromatographic applications was developed for the continuous detection of benzene, toluene, and m-xylene. To improve the resolution of the IMS for single substances a customized IMS with a doubled drift tube length was built. The responses of both IMS (drift tube lengths of 6 and 12 cm) to the compounds selected were determined and compared. The advantages of the combination of multi-capillary columns with IMS are discussed with the aim of achieving further enhancements to the resolution of the instrument. This coupling leads to a significant increase in the scope of application of ion mobility spectrometry for environmental applications. The influence of intermolecular charge-transfer reactions on the peak can be reduced and more complex matrices considered. The presented three-dimensional correlation supports the interpretation of the spectra acquired from the mixtures. © 2000 John Wiley & Sons, Inc. Field Analyt Chem Technol 4: 157–169, 2000
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.