Abstract

Quantitative analysis of amino acids in blood and urine is primarily indicated for the diagnosis of amino acid disorders. The high-performance liquid chromatography (HPLC) technique is frequently used for this detection. The frequency of sample collection on filter paper has been increasing exponentially, and there are many advantages attributed to processing biological samples in this way. The aim of this study was to validate a quantitative analysis of amino acids by HPLC in blood and urine collected on filter paper and to establish reference values in the neonatal period. Dried blood and dried urine samples of respectively 58 and 45 healthy newborns (2-9 days) were collected. Pre-treatment and extraction of samples were done according to the literature. Separation and analysis of amino acids were carried out by HPLC with fluorescence detection. The developed method demonstrated excellent separation, linearity, limits of detection and quantification, repeatability and recovery. The reference values for 17 amino acids were defined in dried blood and urine samples of newborns. This work presents a simple, fast and effective method for the simultaneous analysis of 17 amino acids in blood and urine collected on filter paper in a single run. The reference values were established and validated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.