Abstract

A quantitative analysis of prostate-specific antigen (PSA) in samples of human blood serum by fluorescence immunochromatography using monoclonal antibodies to PSA was developed. The fluorescence immunochromatographic analysis system is composed of anti-PSA-monoclonal antibody (mAb), fluorescence conjugates in detection solution, a immunochromatographic assay strip, and a laser fluorescence scanner. A fluorescence immunochromatographic analysis system was employed to detect PSA on the basis of the area ratio between the control line and the test line of the strip. Under optimal conditions, the area ratio was proportional to PSA concentration ranging from 0.72 to 46.0 ng/mL with a detection limit of 0.72 ng/mL.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.