Abstract
A GC/low-resolution MS method for the quantitative isotope dilution analysis of 39 mono- to heptabrominated diphenyl ethers was developed. The effects of two different ionization sources, electron impact (EI) and electron capture negative ionization (ECNI), and the effects of their parameters on production of high-mass fragment ions [M - xH - yBr](-) specific to PBDEs were investigated. Electron energy, emission current, source temperature, ECNI system pressure, and choice of ECNI reagent gases were optimized. Previously unidentified enhancement of PBDE high-mass fragment ion [M - xH - yBr](-) abundance was achieved. Electron energy had the largest impact on PBDE high-mass fragment ion abundance for both the ECNI and EI sources. By monitoring high-mass fragment ions of PBDEs under optimized ECNI source conditions, quantitative isotope dilution analysis of 39 PBDEs was conducted using nine (13)C(12) labeled PBDEs on a low-resolution MS with low picogram to femtogram instrument detection limits.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.