Abstract

Changes in the proportions of individual isoforms of the MM isoenzyme of creatine kinase (CK; EC 2.7.3.2) in plasma promptly reflect both myocardial infarction and coronary recanalization. However, quantitative methods developed thus far are too slow or cumbersome for routine use in making clinical decisions. We report a convenient, quantitative chromatofocusing assay with on-line fluorometric detection of isoform activity in the column eluent that provides results within 40 min from the time of sample application. Sample eluted from a microbore chromatofocusing column (1.8-mL bed volume) is split between a reaction stream, into which CK reagents are added, and a reference stream. After incubation at 37 degrees C, NADPH formed by reaction of isoforms with CK reagent is detected at 340 nm. The system can detect activity of individual isoforms in plasma samples having total CK activity greater than or equal to 21 U/L (30 degrees C). Results correlated closely with those obtained by previously validated, but slow, chromatofocusing (r = 0.98, n = 30) and protein immunoblotting (r = 0.90, n = 20) procedures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.