IR-spectrometry with full spectra evaluation by means of least squares fitting has been used for the simultaneous quantitative determination of up to 14 organic components at the ppm level in air. The method has been applied to air emissions from chemical production facilities. Tedlar bags have been used for sample collection and transport. Almost any compound with at least weak IR-absorption bands and with a vapor pressure permitting accurate recording of reference spectra can be analyzed. Limits of quantitation are at the low ppm level (at 20 m path length), depending on the absorptivities of the respective compound.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.