Abstract

Amplitude modulation scanning Kelvin probe microscopy (AM-SKPM) is widely used to measure the contact potential difference (CPD) between probe and samples in ambient or dry inert atmosphere. However, AM-SKPM is generally considered quantitatively inaccurate due to crosstalk between the cantilever and the sample. Here we demonstrate that the accuracy of AM-SKPM-based CPD measurements is drastically improved by exciting the SKPM probe at its second eigenmode. In the second eigenmode of oscillation, there exists a stationary node at the cantilever towards its free end, across which the displacement bears opposite signs; therefore driving the SKPM probe at its second eigenmode helps to partially cancel the virtual work done by the cantilever and reduce the crosstalk effect. The improvement in accuracy is experimentally confirmed with interdigitating electrodes calibration samples as well as practical samples such as the cross-section of wafer-bonded GaAs/GaN heterojunction.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call