Abstract

A simple, highly sensitive and specific method based on high-performance liquid chromatography (HPLC) with ultraviolet detection was developed for the measurement of venetoclax concentrations in plasma samples. The chromatographic method employed a mobile phase of acetonitrile: 0.5% KH2PO4 (pH3.5) (80/20, v/v) on a CAPCELL PAK C18 UG120 column at a flow rate of 0.5mL/min. The quantitative method was validated based on standards described in "Bioanalytical Method Validation: Guidance for Industry" published by the US Food and Drug Administration. The separation of venetoclax and the internal standard R051012 was satisfactory, and the chromatograms were free of interfering peaks from the biological matrix. The intra- and inter-day coefficients of variation for venetoclax assays were <12.9%, whereas intra- and inter-day accuracies were within 13.6%. Only 100μL of human plasma was required to detect a lower limit of quantification of 10ng/mL for venetoclax. The recoveries of venetoclax extracted with an Oasis HLB cartridge were between 81 and 85%. The developed HPLC method was successfully applied to the determination of venetoclax concentrations in plasma of acute myeloid leukemia patients taking venetoclax. The degree of drug interactions between venetoclax and CYP3A4 inhibitors can be determined by this HPLC assay.