Abstract

In this investigation, a quantitative matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOFMS) method was developed for the analysis of underivatized free amino acids in mammalian cell culture media. Calibration curves were developed for 12 amino acids over the linear range of 1-100 microM with coefficients of determination ranging from r2 = 0.9220 to r2 = 0.9973. An aerospray method was utilized for the sample deposition method, and the matrix, alpha-cyano-4-hydroxycinnamic acid, served as the internal standard. This assay was used to analyze bioreactor samples from five time points in the process. Concentrations determined through interpolation of the calibration curves were comparable to those obtained via reversed-phase HPLC based analysis with an average percent difference of 19.71%. Repeatability and intermediate precision studies were also performed, and the relative standard deviations ranged from 0.5943 to 21.41 and 3.157 to 18.97, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.