Abstract

Instrumentation and methodology developed for quantitation of stable isotopic traces in urine are described. Calcium is isolated from urine as the insoluble oxalate salt which is subsequently dissolved in hydrochloric acid. The isotopic content of the acid solution is determined by use of a conventional mass spectrometer equipped with a fast atom bombardment ion source. Calcium ions are desorbed from the sample surface by a beam of high-energy xenon atoms and detected with a high-resolution mass spectrometer. A data acquisition system has been developed to control the mass spectrometer and record the ion signals. Detailed analysis of potential sources of error indicates that the precision of the method is presently limited primarily by an isotope effect that occurs during ion desorption. Results presented here demonstrate that the relative abundances of calcium isotopes in urine can be determined with high precision (coefficient of variation < 0.2%) and that the method is a viable alternative to conventional thermal ionization mass spectrometry. The method is especially attractive because it uses a conventional high-resolution mass spectrometer which is routinely used for analysis of organic substances.