Abstract

The effect of coeluting matrix compounds on the quantitation of SR 27417 in human plasma using electrospray liquid chromatography-tandem mass spectrometry has been examined. During the method development stage of this assay, plasma samples spiked with the analyte at 100 pg/mL were extracted using three different procedures: a hexane liquid-liquid extraction, an ethyl acetate back-extraction, and a solid phase extraction. Ion intensity of the analyte was found to be related inversely to the percent ionization of coeluting matrix components as evidenced by full scan spectra. The ethyl acetate back-extraction, which contained the fewest coeluting components, resulted in the highest ion intensity for the analyte. An assay comparison was done by using the liquid-liquid hexane and the ethyl acetate back-extractions for sample preparation. Replicate 1-mL samples (n=5) at 11 concentrations from 5 to 2000 pg/mL were extracted and analyzed. The results for the ethyl acetate back-extracted samples were acceptable from 2000 to 5 pg/mL with accuracy ranging from -11.6 to 2.61% of the nominal concentrations. In contrast, the hexane liquid-liquid method had poor accuracy and precision below 20 pg/mL. The difference is explained by suppression of analyte ion intensity. These results are consistent with the current theory of electrospray ionization.