Abstract

Although levetiracetam (LEV) has favorable linear pharmacokinetic properties, therapeutic drug monitoring (TDM) is necessary for pregnant women with epilepsy. This study aims to build a simple, reliable, and sensitive ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method for determining LEV concentrations in plasma and saliva samples, to support the routine TDM of LEV in Chinese pregnant women with epilepsy. The stable isotope-labeled LEV-d6 was used as the internal standard. The extracted samples were analyzed using a UPLC-MS/MS system with positive electrospray ionization. Mobile phase A was water containing 5 mM ammonium acetate and 0.1% formic acid, and phase B was 1:1 methanol-acetonitrile with 0.1% formic acid. The method was validated and utilized to determine LEV concentrations in non-pregnant and pregnant patients with epilepsy. The developed method was validated in both plasma and saliva samples over a concentration range of 0.1-50 μg/mL. The intra- and inter-batch accuracy for LEV ranged from -7.0% to 2.9%, with precisions between 2.7% and 9.3%. In pregnant patients, the mean dose-standardized LEV trough plasma concentrations were significantly lower than those in non-pregnant patients (4.73 ± 2.99 vs. 7.74 ± 3.59 ng/mL per mg/day; P < 0.0001). It is recommended that the TDM of LEV should be routinely performed during the different stages of pregnancy.