Abstract
Monitoring infliximab (IFX) concentrations and the presence of antidrug antibodies (ADA) is important for patient management. We developed a method to measure IFX and ADA in serum in a single injection using surface plasmon resonance (SPR). Using the Bio-Rad ProteOn XPR36, tumor necrosis factor-α and IFX were covalently immobilized onto separate lanes of a chip surface. Diluted serum was injected over both lanes, followed by an injection of goat antihuman antibody. The binding response was used to quantify IFX or detect ADA. The analytical performance of the assay was determined. Using 50 patient samples, SPR results were compared with results from a reporter gene assay (RGA). For the quantification of IFX, the functional sensitivity was 0.5 μg/mL. The total precision was <10% for all concentrations tested. IFX concentrations measured by SPR correlated well with RGA (R = 0.862), but a bias was observed (slope = 0.61). SPR detected 14 ADA-positive samples. Compared with RGA for ADA detection, there were 6 true-positive, 8 false-positive, 5 false-negative, and 31 true-negative findings. SPR can be used to measure biological drug concentrations and detect ADA in serum. This technique may provide complementary information to current methods used to detect ADA.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.