Abstract

Real-time PCR technology may provide an accurate and sensitive method to quantify hepatitis C virus (HCV) RNA. So far, studies have been carried out using the Taqman technology with the ABI Prism 7700 sequence detector. An alternative and simple real-time PCR assay is described with no probe requirement, based on the SYBR Green I dye and LightCycler™ fluorimeter. Amplicon synthesis was monitored continuously by SYBR Green I dye binding to double stranded DNA during PCR of the 5′ HCV non-coding (NC) region. Specificity was verified by amplicon melting temperatures. An external standard curve was constructed with serial 10 fold dilutions of a modified synthetic HCV 5′ NC RNA. A wide range linear relationship (up to 3.7×10 9 copies/ml) was observed between number of PCR cycle needed to detect a fluorescent signal and number of RNA copy. Intra- and inter-assay coefficients of variation were 0.7 to 2.1 and 3.7% respectively, indicating good reproducibility of the method. Thirty-three HCV positive sera of different genotypes were quantified by this method and gave similar but more sensitive results compared to the branched DNA (bDNA) technology.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.