Ethyl glucuronide and ethyl sulfate are minor conjugated metabolites of ethanol that can be detected in urine for several days after last ingestion of ethanol. The monitoring of ethanol use has both clinical and forensic applications and a longer detection window afforded by monitoring these metabolites is obvious. LC-MS/MS is used to analyze diluted urine with deuterated analogs of each analyte as internal standards to ensure accurate quantitation and control for any potential matrix effects. High aqueous HPLC is used to chromatograph the metabolites. Negative ion electrospray is used to introduce the metabolites into the mass spectrometer. Selected reaction monitoring of two product ions for each analyte allows for the calculation of ion ratios which ensures correct identification of each metabolite, while a matrix-matched calibration curve is used for quantitation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.