Abstract

A simple solid phase extraction (SPE) technique combined with gas chromatography-mass spectrometry (GC/MS) operated in selected ion monitoring (SIM) mode is described for quantitation of cocaine and cocaethylene in small samples (250 μl) of rat whole blood. Use of ( N-[ 2H 3C])-cocaine and (N- [ 2H 3C])-cocaethylene internal standards resulted in high sensitivity and selectivity for this analytical method. Analysis was performed using a Hewlett-Packard 5890 GC equipped with a 7673A Automatic Liquid Sampler linked to a Hewlett-Packard 5972 Mass Selective Detector. Separation of analytes was accomplished on a cross-linked methyl silicone gum capillary column (Ultra 1: 12m × 0.2mm (i.d.) × 0.33 μm). Linearity was established over a wide range of concentrations (5.0–2000.0 ng ml −1) with good precision. Limits of detection (LOD) were 1.0 and 2.0 ng ml −1 for cocaine and cocaethylene, respectively. This analytical method was designed for use in pharmacokinetic experiments studying the formation of cocaethylene following ethanol pretreatment in rats administered cocaine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.