Abstract

We measured beta-lipoprotein in 232 cord sera by rate nephelometry and compared the results with those obtained by electroimmunoassay. The mean (mean +/- two standard deviations) was 544 (293-1009) mg/L by rate nephelometry and 503 (265-955) mg/L by electroimmunoassay. The between-assay correlation coefficient was 0.891. Between-run coefficients of variation were 3.5-4.4% for nephelometry and 11-14% for electroimmunoassay. Nephelometry was faster, more precise, and less labor intensive than electroimmunoassay. For these reasons, rate nephelometry appears to be a more suitable assay system for mass screening, should detection of familial hypercholesterolemia prove feasible and desirable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.