Abstract

We developed and validated an analytical method for quantifying 25-hydroxyvitamin D2 (25OHD2) and 25-hydroxyvitamin D3 (25OHD3) in serum and plasma. Samples, pretreated with zinc sulfate and methanol, were extracted with hexane. Separation was achieved via UHPLC and 25OHD quantification was accomplished by a triple quadrupole mass spectrometer. Imprecision was 3.6-15.1%CV and bias 88.0-126.0%. Extraction efficiency was 76.5-110.5%, whereas the matrix effect ranged from -46.7 to -32.0%. The method was applied to authentic specimens. The results showed no significant difference between serum and plasma; strong correlation with paired values from an external laboratory; and analyte stability for 15 days. This method provides reliable and accurate measurement of 25OHD for use in clinical practice.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call