Abstract
We developed and validated an analytical method for quantifying 25-hydroxyvitamin D2 (25OHD2) and 25-hydroxyvitamin D3 (25OHD3) in serum and plasma. Samples, pretreated with zinc sulfate and methanol, were extracted with hexane. Separation was achieved via UHPLC and 25OHD quantification was accomplished by a triple quadrupole mass spectrometer. Imprecision was 3.6-15.1%CV and bias 88.0-126.0%. Extraction efficiency was 76.5-110.5%, whereas the matrix effect ranged from -46.7 to -32.0%. The method was applied to authentic specimens. The results showed no significant difference between serum and plasma; strong correlation with paired values from an external laboratory; and analyte stability for 15 days. This method provides reliable and accurate measurement of 25OHD for use in clinical practice.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.