Abstract

This paper presents the evaluation of the geometry correction factor for calibration of Bonner sphere spectrometer (BSS) equipped with the 3He counter (typical SP9), based on Monte Carlo simulation. Using the MCNP4C code, geometry factors and effectiveness parameters (δ) of a 8″ polyethylene sphere (r = 10.16 cm) were calculated for 10 different energies and 3 various distances between the source and center of the sphere. The obtained results showed that the geometry factor increases with the distance, and the effectiveness parameter is independent of distance. Finally, for 2″, 3″, 5″, 8″ and 12″ spheres exposed to four different radio-isotopic neutron sources with various energy spectrums, the effectiveness parameters were determined, which it is worthy to consider δ ≡ δ(r, E).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.