Abstract

Methionine is a common excipient used in therapeutic protein liquid formulations as stabilizer and antioxidant. The oxidation of methionine to methionine sulfoxide can be regarded as a sensitive marker of oxidative stress for drug product storage conditions. In this study, a sensitive HPLC method for the quantification of methionine sulfoxide in formulated protein product was developed and qualified according to regulatory requirements using a SIELC® Primesep 100 column with UV detection. The separation involves a mixed-mode mechanism including reversed phase and cationic exchange modalities. The operating range of the method was established between 1 µM and 35 µM of methionine sulfoxide. In this testing range, the method was shown to be linear (R2 > 0.99), accurate (Recovery 92.9 – 103.6%, average recovery = 99.8 ± 1.4%) and precise (intermediate precision at LoQ, CV = 2.9%). The developed test system was successfully applied to study the effects of temperature and storage conditions on methionine sulfoxide formation in complex therapeutic antibody formulations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.