Abstract

A methodology has been developed to chromatographically quantify indium in polymetallic (bio)hydrometallurgical processing solutions using the Dionex IonPac CS5A column and pyridine-2,6-dicarboxylic acid eluent. Cu(II) and In(III) could be separated by elevating the column temperature to 45°C. The comparatively low stability constant of the In-eluent complex (log K2 =3.8) required typical leaching samples to be diluted in the eluent rather than acid or water to overcome ligand competition between components of the sample solution and the eluent. The methodology was applied to leachates from (bio)hydrometallurgical processing of oxidic flue dust residues and sulfidic zinc ores, where both are promising candidates for the recovery of indium from low grade ores and metallurgical wastes. Indium, ferrous iron, ferric iron, copper, zinc, nickel, and manganese concentrations could be simultaneously quantified. The method was found suitable for samples containing at least 0.25mg/L indium and an iron to indium ratio of up to 100:1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.