Abstract

An analysis method has been developed for determining low parts-per-quadrillion by volume (ppqv) concentrations of nerve agent VX vapor actively sampled from ambient air. The method utilizes glass bead filled depot area air monitoring system (DAAMS) sampling tubes with isopropyl alcohol extraction and isotope dilution using liquid chromatography coupled with a triple-quadrupole mass spectrometer (LC/MS/MS) with positive ion electrospray ionization for quantitation. The dynamic range was from one-tenth of the worker population limit (WPL) to the short-term exposure limit (STEL) for a 24 L air sample taken over a 1 h period. The precision and accuracy of the method were evaluated using liquid-spiked tubes, and the collection characteristics of the DAAMS tubes were assessed by collecting trace level vapor generated in a 1000 L continuous flow chamber. The method described here has significant improvements over currently employed thermal desorption techniques that utilize a silver fluoride pad during sampling to convert VX to a higher volatility G-analogue for gas chromatographic analysis. The benefits of this method are the ability to directly analyze VX with improved selectivity and sensitivity, the injection of a fraction of the extract, quantitation using an isotopically labeled internal standard, and a short instrument cycle time.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.