Abstract

A simple high-performance liquid chromatographic (HPLC) method was developed for the assay of total and free mycophenolic acid (MPA) in human plasma. Prior to analysis, total mycophenolic acid was extracted by protein precipitation and free drug was isolated from plasma samples using ultrafiltration. The extracts were injected onto a Kromasil C 8 column at 30 °C with excitation and emission wavelengths set at 342 and 425 nm, respectively. The mobile phase was consisted of acetonitrile-32 mM glycine buffer, pH 9.2 (20:80, v/v), at a flow rate of 1.0 ml/min. The method was found to be linear over the concentration range investigated, 0.05–40 mg/l for total mycophenolic acid ( r > 0.999) and 5–1000 μg/l ( r > 0.99) for free drug. The percentage error of the analytical method was below 10.9%. The intra- and inter-day reproducibility was adequate with the coefficients of variation of 8.28% or below. The run time were 4 and 6 min for free and total MPA, respectively. The method thus can be effectively applied to measure mycophenolic acid concentrations in clinical samples.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call