Abstract

Occlusal splints are commonly used to position the maxilla during traditional orthognathic surgery. We aimed to quantify the inaccuracy of the maxillary positioning (in three dimensions) in traditional model surgery with the Orthopilot® navigation system. Thirty Le Fort I osteotomies were made using a standard technique. The position of the maxilla was recorded by the navigation system and defined by three values of translation and three of rotation. The recorded data were compared with the planned data. The accuracy of positioning was classified in distinct classes with three major criteria (conformity, non-conformity, and failure) according to the discrepancy. The positioning of the maxilla was in conformity with operative planning in 3/30 of our Le Fort I osteotomies (95% CI 2% to 27%) and in failure in 22/30 (95% CI 54% to 88%). The dispersion of the discrepancy was more important in the sagittal plane, particularly for the sagittal rotation and for the back-front translation, which reflected greater inaccuracy in this plan. The frontal orientation of the maxilla was better controlled. The risk of maxillary malposition was proportional to the planned maxillary advancement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.