Abstract

Measurement of myocardial and brain perfusion when using exogenous contrast agents (CAs) such as gadolinium-DTPA (Gd-DTPA) and MRI is affected by the diffusion of water between compartments. This water exchange may have an impact on signal enhancement, or, equivalently, on the longitudinal relaxation rate, and could therefore cause a systematic error in the calculation of perfusion (F) or the perfusion-related parameter, the unidirectional influx constant over the capillary membranes (K(i)). The aim of this study was to quantify the effect of water exchange on estimated perfusion (F or K(i)) by using a realistic simulation. These results were verified by in vivo studies of the heart and brain in humans. The conclusion is that water exchange between the vascular and extravascular extracellular space has no effect on K(i) estimation in the myocardium when a normal dose of Gd-DTPA is used. Water exchange can have a significant effect on perfusion estimation (F) in the brain when using Gd-DTPA, where it acts as an intravascular contrast agent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.