Abstract
AbstractThis work reports the highly sensitive detection of quercetin using glassy carbon electrodes modified with multiwall carbon nanotubes (MWCNT) dispersed in polyethylenimine and poly(acrylic acid). The adsorptive stripping with square wave voltammetric transduction allowed the detection of 7.5 nM quercetin after 1.0 min accumulation at open circuit potential. The amperometric detection at 0.150 V is reported as an easier and simpler alternative to determine quercetin, with detection limits of 0.2 µM. The careful selection of the working conditions also made possible the detection of submicromolar levels of quercetin in the presence of excess of rutin. The proposed methodology was successfully used to quantify quercetin in onion samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.