Abstract

Polychlorinated dibenzo- p-dioxins and dibenzofurans in crude extracts of fly ash and flue gas from municipal waste incinerators were quantified using a comprehensive multidimensional gas chromatograph (GC × GC) coupled to a high-resolution time-of-flight mass spectrometer (HR-TOFMS). For identification and quantification, we developed our own program to prepare 3D chromatograms of selected mass numbers from the data of the GC × GC/HR-TOFMS. Isolation of all congeners with a TCDD toxic equivalency factor from the other isomers by only one injection was confirmed. The instrumental detection limit of TCDD on the GC × GC/HR-TOFMS was 0.9 pg by the relative calibration method. Quantification of these substances in the crude extracts was achieved by direct injection to the GC × GC/HR-TOFMS. The results agree with the values obtained using a generic gas chromatography/high-resolution mass spectrometry (GC/HRMS) system. It was confirmed that measurement by high-resolution TOFMS and GC × GC effectively reduces interference from other chemicals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.