Abstract

An analytical method for the analysis of relevant secoiridoid-based components in olive oil, oleacein and oleuropein aglycone, is described using for the first time deuterated surrogates. 0.2g of sample was necessary to perform the analysis using liquid-liquid extraction and ultrasound-assisted extraction with a mixture of methanol/water (4:1, v/v). To avoid the formation of by-products, normal-phase ultra high performance liquid chromatography was chosen for the chromatographic separation. The selected mobile phase was a gradient mixture of tetrahydrofurane and hexane, and an ACE Excel 3 CN-ES column as stationary phase. The detection and quantification was performed with a SYNAPT G2-Si mass spectrometer. The calibration curves for oleacein and oleuropein aglycone were linear and quadratic, respectively. The validation was done at three levels of concentration. Relative errors from 0.1 to 10.5% and relative standard deviations lower than 9% were obtained. The method was applied to study different samples of olive oil.