Abstract

Nepetalactones, the major chemical components of catnip (Nepeta cataria L.), were analysed by reversed-phase HPLC coupled with UV and MS detection. Two major nepetalactones, Z,E-nepetalactone and E,Z-nepetalactone, were successfully identified and quantified. The linearity range for Z,E-nepetalactone was determined as 0.00655-0.655 mg/mL with a correlation coefficient of 0.9999, and the linearity range of E,Z-nepetalactone was found to be 0.00228-0.456 mg/mL with a correlation coefficient of 0.9999, under UV detection at 228 nm. The linearity ranges were from 0.00164 to 0.0328 mg/mL, with a correlation coefficient of 0.9999, for ZE-nepetalactone and 0.00114-0.0228 mg/mL, with a correlation coefficient of 0.9999, for E,Z-nepetalactone by MS detection with selected ion monitoring of ion peak m/z 167. The MS detection was found to be more sensitive than UV detection and this method was validated as simple, reliable and sensitive for catnip nepetalactone analysis. This method can be used for identification and fingerprinting of catnip products.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.