Abstract
Mycophenolate mofetil (MMF) and sodium mycophenolate are commonly prescribed immunosuppressive drugs for patients who have undergone solid organ transplant. Therapeutic drug monitoring (TDM) of these drugs is performed by assessing mycophenolic acid (MPA) in plasma. Due to the large inter-individual variability and narrow therapeutic range, the precise determination of systemic MPA concentration carries great clinical significance. We present a rapid, sensitive, specific, and robust liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the quantitation of MPA in plasma. A Waters Xevo TQ-S Micro mass spectrometer coupled to a Water's Acquity liquid chromatography system was used in positive electrospray ionization (ESI) mode. MPA quantitation was achieved using multiple reaction monitoring (MRM). Mycophenolic acid carboxybutoxy ether (MPAC) was employed as an internal standard. The method is linear from 0.25 to 40.00mg/L, has intra-assay (N=24) imprecision of 2.7% at 1.57mg/L and 3.9% at 4.61mg/L and inter-assay (N=20days) imprecision of 4.0% at 1.62mg/L and 5.6% at 4.68mg/L.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.