Abstract

The aim of this study was to validate a novel method of determining vena contracta area (VCA) and quantifying mitral regurgitation using multibeam high-pulse repetition frequency (HPRF) color Doppler. The Doppler signal was isolated from the regurgitant jet, and VCA was found by summing the Doppler power from multiple beams within the vena contracta region, where calibration was done with a reference beam. In 27 patients, regurgitant volume was calculated as the product of VCA and the velocity-time integral of the regurgitant jet, measured by continuous-wave Doppler, and compared with regurgitant volume measured by magnetic resonance imaging (MRI). Spearman's rank correlation and the 95% limits of agreement between regurgitant volume measured by MRI and by multibeam HPRF color Doppler were r(s) = 0.82 and -3.0 +/- 26.2 mL, respectively. For moderate to severe mitral regurgitation, there was good agreement between MRI and multibeam HPRF color Doppler. Agreement was lower in mild regurgitation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.